SPECIFICATIONS
ELECTRICAL
Temperature Scale: Celsius or Fahrenheit user-selectable
Measurement Range: -50°C to 1300°C, (-58°F to 2000°F)
Resolution: 1°C or 1°F, 0.1°C or 0.1°F
Accuracy: Accuracy is specified for operating temperatures over the range of
18°C to 28°C (64°F to 82°F), for 1 year, not including thermocouple error.
±(0.3% rdg + 1°C) -50°C to 1000°C
±(0.5% rdg + 1°C) 1000°C to 1300°C
±(0.3% rdg + 2°F) -58°F to 2000°F
Temperature Coefficient: 0.1 times the applicable accuracy specification
per °C from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and 82°F to 122°F).
Input Protection: 24V dc or 24V ac rms maximum input voltage on any
combination of input pins.
Reading Rate: 2.5 times per second.
Input Connector: Accepts standard miniature thermocouple connectors (flat
blades spaced 7.9mm, center to center).